26582 CRRT circuit HEPUF

CRRT circuit HEPUF
Test Code: HEPUF-CIR
Synonyms/Keywords

circuit, ECMO, HEPUF, anti-Xa​

Test Components

​HEPUF​

Specimen Requirements
Fasting RequiredSpecimen TypePreferred Container/TubeAcceptable Container/TubeSpecimen VolumeSpecimen Minimum Volume
(allows for 1 repeat)
Pediatric Minimum Volume
(no repeat)
​No
​Plasma
​3.2% Citrated Blue Top Tube (BTT)


​1 mL
​0.6 mL
Collection Processing Instructions

​MMC-Marshfield ONLY

ECMO patients: Citrated Blue Top Tube (BTT) must be drawn to the fill line. Completely invert tube 3 to 4 times to mix. Do Not Shake! 

Specimen Stability Information
Specimen TypeTemperatureTime
​Plasma
​Ambient
​4 hours
Rejection Criteria

Gross Hemolysis

Specimens filled below the fill line

Clotted

Performing Laboratory Information
Performing LocationDay(s) Test PerformedAnalytical TimeMethodology/Instrumentation
​Marshfield
​Monday through Sunday
​Less than 2 hours
​Anti-Factor Xa Chromogenic/Sysmex CS-2500
Reference Range Information

No reference range established for this analyte/sample type

Synonyms/Keywords

circuit, ECMO, HEPUF, anti-Xa​

Test Components

​HEPUF​

Ordering Applications
Ordering ApplicationDescription​
​Cerner
​CRRT circuit HEPUF
If the ordering application you are looking for is not listed, contact your local laboratory for assistance.
Specimen Requirements
Fasting RequiredSpecimen TypePreferred Container/TubeAcceptable Container/TubeSpecimen VolumeSpecimen Minimum Volume
(allows for 1 repeat)
Pediatric Minimum Volume
(no repeat)
​No
​Plasma
​3.2% Citrated Blue Top Tube (BTT)


​1 mL
​0.6 mL
Collection Processing

​MMC-Marshfield ONLY

ECMO patients: Citrated Blue Top Tube (BTT) must be drawn to the fill line. Completely invert tube 3 to 4 times to mix. Do Not Shake! 

Specimen Stability Information
Specimen TypeTemperatureTime
​Plasma
​Ambient
​4 hours
Rejection Criteria

Gross Hemolysis

Specimens filled below the fill line

Clotted

Test Components

​HEPUF​

Reference Range Information

No reference range established for this analyte/sample type

For more information visit:
Performing Laboratory Information
Performing LocationDay(s) Test PerformedAnalytical TimeMethodology/Instrumentation
​Marshfield
​Monday through Sunday
​Less than 2 hours
​Anti-Factor Xa Chromogenic/Sysmex CS-2500
For billing questions, see Contacts
For most current information refer to the Marshfield Laboratory online reference manual.